M4000 Insulation Analyzer User Guide
72A-1230 Rev. F 4-27
July 22, 2005
Figure 4.4 Calibration Verification Initial Screen
2. Press F2.
3. Follow the on-screen directions, as shown in Figure 4.5.
Figure 4.5 Calibration Verification Instructions
Connections for the Calibration Verification are shown in Figure 4.6 and
Figure 4.7. The Field Adapter cable, shown in these figures, is an
accessory cable found in the cable bag which connects from the High
Voltage Test Cable to the black connectors on the Field Calibration
Reference.